Pressure/Vacuum

Bob Quinn completes an air velocity measurement        Bob Quinn performs a mass flow calibration

ORNL Metrology Laboratory (ORNLML) maintains a variety of pressure and vacuum standards traceable to SI (International System of Units) through the National Institute of Standards and Technology (NIST) and other international primary laboratories. These standards generate and measure various pressure and vacuum properties required to perform traceable calibrations of our customers’ instruments and equipment.  ORNLML can provide accredited calibrations for any pressure parameters within the ranges detailed in our Scope of Accreditation. However, traceable non-accredited calibrations are available for pressure and vacuum outside of our Scope of Accreditation. The tables below list our accredited capabilities. Our entire Scope of Accreditation can be found at the following link.

Typical types of equipment the ORNLML performs pressure and vacuum calibrations on are: Pressure gauges and transducers, barometers, manometers, capacitance gauges and vacuum gauges.

If you have questions about our pressure and vacuum capabilities, please contact the ORNL Metrology Laboratory for more information.

Capability
Pressure –Absolute Mode (20/T05)


Range

Best Uncertainty (+/-) in ppm + mPa (psi) *

Remarks

1.0 Pa to 15.0 kPa

37 + 8

Source/Measure

0.0000145 psia to 2.17 psia

37 + 1.2 E-6

Source/Measure

 

Range

Best Uncertainty (+/-) in ppm of Reading *

Remarks

13.8 kPa to 34.5 kPa

98.3

Source/Measure

2.0 psia to 5 psia

98.3

Source/Measure

 

 

 

34.5 kPa to 172.4 kPa

43.2

Source/Measure

5 psia to 25 psia

43.2

Source/Measure

 

 

 

82.7 kPa to 3447.4 kPa

25.2

Source/Measure

12 psia to 500 psia

25.2

Source/Measure

 

 

 

1665 kPa to 6894.8 kPa

21.4

Source/Measure

24 psia to 1000 psia

21.4

Source/Measure

Absolute Differential Mode

Range

Best Uncertainty (+/-) in ppm + mPa (psi) *

Remarks

0 Pa to 15.0 kPa

37 + 5

Source/Measure

0 psia to 2.17 psia

37 + 7.3 E-7

Source/Measure

Gauge Mode

Range

Best Uncertainty (+/-) in ppm + mPa (psi) *

Remarks

0 Pa to 15.0 kPa

37 + 5

Source/Measure

0 psi to 2.17 psi

37 + 7.3 E-7

Source/Measure

 

Range

Best Uncertainty (+/-) in ppm of Reading *

Remarks

5.5 kPa to 172.4 kPa

19.3

Source/Measure

0.8 psig to 25 psig

19.3

Source/Measure

 

 

 

82.7 kPa to 6894 kPa

19.8

Source/Measure

5 psig to 25 psig

19.8

Source/Measure

 

 

 

1.379 MPa to 68.948 MPa

108

Source/Measure

200 psig to 10000 psig

108

Source/Measure

 

 

 

1665 kPa to 6894.8 kPa

21.4

Source/Measure

24 psia to 1000 psia

21.4

Source/Measure

*Represents an expanded uncertainty using a coverage factor, k=2, at an approximate level of confidence of 95%.

 