FDA Home Page | CDRH Home Page | Search | A-Z Index | ||
Information
for mammography facility personnel, inspectors, and consumers about the implementation of the Mammography Quality Standards Act of 1992 (MQSA) |
1. Conducting the daily processor QC tests when the sensitometer is not availableThis alternative standard was approved on October 18, 1999 and was made retroactive to April 28, 1999. It has no time limit. The alternative to sensitometric-densitometric testing of processor performance can be used for a period of up to two weeks when the facility's sensitometer is unavailable. This alternative is based on evaluating a phantom image through measurements described in 21 CFR 900.12(e)(1) and (2). The final regulation and its alternative standard are stated below: 21 CFR 900.12(e)(1) and (2) states that: |
||||||||||||||||||||||||||||||||||||||||||||||||
(1) | Daily quality control tests. Film processors used to develop mammograms shall be adjusted and maintained to meet the technical development specifications for the mammography film in use. A processor performance test shall be performed on each day that clinical films are processed before any clinical films are processed that day. The test shall include an assessment of base plus fog density, mid-density, and density difference, using the mammography film used clinically at the facility. | |||||||||||||||||||||||||||||||||||||||||||||||
(i) | The base plus fog density shall be within + 0.03 of the established operating level. | |||||||||||||||||||||||||||||||||||||||||||||||
(ii) | The mid-density shall be within +/- 0.15 of the established operating level. | |||||||||||||||||||||||||||||||||||||||||||||||
(iii) | The density difference shall be within +/- 0.15 of the established operating level. | |||||||||||||||||||||||||||||||||||||||||||||||
(2) | Weekly quality control tests. Facilities with screen-film systems shall perform an image quality evaluation test, using an FDA-approved phantom, at least weekly. | |||||||||||||||||||||||||||||||||||||||||||||||
(i) | The optical density of the film at the center of an image of a standard FDA-accepted phantom shall be at least 1.20 when exposed under a typical clinical condition. | |||||||||||||||||||||||||||||||||||||||||||||||
(ii) | The optical density of the film at the center of the phantom image shall not change by more than +/- 0.20 from the established operating level. | |||||||||||||||||||||||||||||||||||||||||||||||
(iii) | The phantom image shall achieve at least the minimum score established by the accreditation body and accepted by FDA in accordance with Sec. 900.3(d) or Sec. 900.4(a)(8). | |||||||||||||||||||||||||||||||||||||||||||||||
(iv) | The density difference between the
background of the phantom and an added test object, used to assess image
contrast, shall be measured and shall not vary by more than +/- 0.05 from
the established operating level. |
|||||||||||||||||||||||||||||||||||||||||||||||
When using the alternative test, processor
performance is considered satisfactory if: |
||||||||||||||||||||||||||||||||||||||||||||||||
1. | The optical density of the film at the center of an image of a standard FDA-accepted phantom is at least 1.20 when exposed under typical clinical conditions. | |||||||||||||||||||||||||||||||||||||||||||||||
2. | The optical density of the film at the center of the phantom image changes no more than +/- 0.20 from the established operating level. | |||||||||||||||||||||||||||||||||||||||||||||||
3. | The density difference between the
background of the phantom and an added test object, used to assess image
contrast, is measured and does not vary by more than +/-0.05 from the established
operating level. |
|||||||||||||||||||||||||||||||||||||||||||||||
In addition: |
||||||||||||||||||||||||||||||||||||||||||||||||
4. | To evaluate base + fog, an additional
measurement of density must be made either of a shielded portion of the
phantom image film or of an unexposed film. In accordance with 21 CFR 900.12(e)(1)(i),
the base plus fog density must be within + 0.03 of the established operating
level. |
|||||||||||||||||||||||||||||||||||||||||||||||
This alternative test must be conducted "each day clinical films are processed, but before processing of clinical films." All results must be recorded and charted. If processor performance fails to meet any part of the alternative test, the problem must be corrected before processing is resumed. 2. Continuous display of the override status for machines with decompression devicesThis alternative standard was approved on June 22, 1999 and was made retroactive to April 28, 1999. It has no time limit. The final regulation and its alternative standard are stated below: 21 CFR 900.12(e)(5)(xi) states that: |
||||||||||||||||||||||||||||||||||||||||||||||||
(xi) | Decompression. If the system is equipped with a provision for automatic decompression after completion of an exposure or interruption of power to the system, the system shall be tested to confirm that it provides: | |||||||||||||||||||||||||||||||||||||||||||||||
(A) | An override capability to allow maintenance of compression; | |||||||||||||||||||||||||||||||||||||||||||||||
(B) | A continuous display of the override status; and | |||||||||||||||||||||||||||||||||||||||||||||||
(C) | A manual emergency compression release
that can be activated in the event of power or automatic release failure. |
|||||||||||||||||||||||||||||||||||||||||||||||
The approved alternative standard to
21 CFR 900.12(e)(5)(xi)(B) allows facilities having machines equipped with
automatic decompression devices that are never disabled to permanently place
a label on the panel indicating that the unit must always be operated in
the automatic decompression mode, in lieu of a continuous display of the
automatic decompression override status required in 21 CFR 900.12(e)(5)(xi)(B).
The wording of this label must be: |
||||||||||||||||||||||||||||||||||||||||||||||||
"Unit always to be used in auto release
mode. If auto release is overridden this status will not be displayed." |
||||||||||||||||||||||||||||||||||||||||||||||||
3. Conducting the weekly phantom image test at facilities with intermittent mammography operationThis alternative standard was approved on May 24, 1999 and was made retroactive to April 28, 1999. It has no time limit. It applies to facilities that do not conduct mammography every week. Rather, they may conduct mammography during some, but not all, weeks in a given month. The final regulation and its alternative standard are stated below: 21 CFR 900.12(e)(2) states that: |
||||||||||||||||||||||||||||||||||||||||||||||||
(2) | Weekly Quality Control Tests. Facilities
with screen-film systems shall perform an image quality evaluation test,
using a FDA-approved phantom, at least weekly. |
|||||||||||||||||||||||||||||||||||||||||||||||
The approved alternative standard is: |
||||||||||||||||||||||||||||||||||||||||||||||||
(2) | Weekly Quality Control Tests. Facilities
with screen-film systems shall perform an image quality evaluation test,
using an FDA-approved phantom, in each week that clinical mammography examinations
are performed, prior to the performance of such examinations. |
|||||||||||||||||||||||||||||||||||||||||||||||
The alternative standard requires that if the number of weeks per month in which clinical mammography is performed increases or decreases, the frequency of the performance of the phantom image quality test must automatically undergo a corresponding increase or decrease. 4. Post exposure indication of the machine pre-selected focal spot and or target materialThis alternative standard was approved on April 19, 1999 and became effective on April 28, 1999 for Senographe™ DMR GE machines. It has no time limit. The final regulation and its alternative standard are stated below: 21 CFR 900.12(b)(7) states that: |
||||||||||||||||||||||||||||||||||||||||||||||||
(7) | Focal spot selection. | |||||||||||||||||||||||||||||||||||||||||||||||
(i) | When more than one focal spot is provided, the system shall indicate, prior to exposure, which focal spot is selected. | |||||||||||||||||||||||||||||||||||||||||||||||
(ii) | When more than one target material is provided, the system shall indicate, prior to exposure, the preselected target material. | |||||||||||||||||||||||||||||||||||||||||||||||
(iii) | When the target material and/or
focal spot is selected by a system algorithm that is based on the exposure
or on a test exposure, the system shall display, after the exposure, the
target material and/or focal spot actually used during the exposure. |
|||||||||||||||||||||||||||||||||||||||||||||||
The approved alternative is: |
||||||||||||||||||||||||||||||||||||||||||||||||
(7) | Focal spot selection. | |||||||||||||||||||||||||||||||||||||||||||||||
(i) | When more than one focal spot and/or more than one target material is provided, the system shall indicate, prior to exposure, the pre-selected focal spot and target material, and shall indicate, after the exposure, the focal spot and test material actually used during the exposure; or | |||||||||||||||||||||||||||||||||||||||||||||||
(ii) | When the target material and/or focal
spot is selected by a system algorithm that is based on the exposure or
on a test exposure, the system shall indicate, after the exposure, the target
material and/or focal spot actually used during the exposure. |
|||||||||||||||||||||||||||||||||||||||||||||||
Under the approved alternative, an indication of the pre-exposure focal spot and target material would no longer be required when the pre-exposure target material and focal spot are set by a system algorithm based on exposure and the user has no control over that selection. In operating modes where the user has control of the pre-selected focal spot and/or target material, indication of the pre-selected values would still be required. In all cases, indication of the focal spot and/or target material actually used during the exposure would be required. 5. Verification Testing After Certain Modifications of the AEC of Senographe™ 700T, 800T, DMR Mammography SystemsThis alternative standard was approved and became effective on September 24, 2001. It has no time limit. Under this alternative, the verification after the specified AEC modifications may be performed under the conditions of Medical Physicist Oversight. The final regulation and its alternative standard are stated below: 21 CFR 900.12(e)(10) states that: |
||||||||||||||||||||||||||||||||||||||||||||||||
(10) | Mammography equipment evaluations.
Additional evaluations of mammography units or image processors shall be
conducted whenever a unit or processor is installed, a unit or processor
is dissembled and reassembled at the same or a new location, or major components
of a mammography unit or processor equipment are changed or repaired. These
evaluations shall be used to determine whether the new or changed equipment
meets the requirements of applicable standards in paragraphs (b) and (e)
of this section. All problems shall be corrected before the new or changed
equipment is put into service for examinations or film processing. The mammography
equipment evaluation shall be performed by a medical physicist or by an
individual under the direct supervision of a medical physicist. |
|||||||||||||||||||||||||||||||||||||||||||||||
The alternative applies to the specific situations described below, in which manufacturer’s software modifications have been made to specific units. In such cases, an onsite mammography equipment evaluation performed by the medical physicist is not required. Instead, all that is required is oversight by the medical physicist. The approved alternative standard
states that for: |
||||||||||||||||||||||||||||||||||||||||||||||||
1. | the modification of the AEC component of Senographe™ 700T or 800T mammography systems described in the GE Medical System's Field Modification Instruction (FMI) 11451, "Seno 700/800T Optical Density Optimization", and | |||||||||||||||||||||||||||||||||||||||||||||||
2. | the optimization of the AEC component of the Senographe™ DMR mammography systems described in the GE Medical System's FMI 11450, "CMR V1/V2+ Optical Density Optimization": | |||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||||
6. Conducting the Mammography Equipment Evaluation After a Software Upgrade Under Medical Physicist OversightThis amended alternative standard was originally approved and became effective on May 31, 2002. It was last amended on October 26, 2007. It has no time limit. It defines the conditions under which the mammography equipment evaluations performed after some computer software upgrades may be performed either by a medical physicist on site or under the conditions of Medical Physicist Oversight. If these conditions are not met, the mammography equipment evaluation after the upgrade must be performed by a medical physicist on site. The original standard is contained within 21 CFR 900.12(e)(10): |
||||||||||||||||||||||||||||||||||||||||||||||||
(10) | Mammography equipment evaluations. Additional evaluations of mammography units or image processors shall be conducted whenever a unit or processor is installed, a unit or processor is dissembled and reassembled at the same or a new location, or major components of a mammography unit or processor equipment are changed or repaired. These evaluations shall be used to determine whether the new or changed equipment meets the requirements of applicable standards in paragraphs (b) and (e) of this section. All problems shall be corrected before the new or changed equipment is put into service for examinations or film processing. The mammography equipment evaluation shall be performed by a medical physicist or by an individual under the direct supervision of a medical physicist. |
|||||||||||||||||||||||||||||||||||||||||||||||
The approved alternative and the conditions for its use are:
Additional conditions for using this alternative requirement in association with a software upgrade are that
If all the above (1-8) conditions are met, then the mammography equipment evaluation may be conducted either during an onsite visit by a medical physicist or under Medical Physicist Oversight. If any of the necessary tests after the software upgrade are required to be performed by the medical physicist, the mammography equipment evaluation must be performed in its entirety by the medical physicist on site. FDA’s acknowledgement as stated in item 7 above does not constitute approval of the type of testing of the software upgrade in question. FDA will contact the FFDM manufacturer if any issues arise concerning the upgrade in question. As a result, FDA may disallow testing the software upgrade under Medical Physicist Oversight if it determines that one or more of the above conditions were not met. Before the October 26, 2007 amendment which generalized the conditions for approval, General Electric (GE) applied to the FDA and obtained approval to conduct the mammography equipment evaluation (MEE) under Medical Physicist Oversight for the software upgrades and to the GE units shown in the Table below:
|
||||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||||
Notification of Software Upgrades Under Amended Alternative Standard #6The new amendments provide a revised FDA approval process for new software upgrades that eliminates the need for FDA to specifically approve each new software upgrade for which mammography equipment evaluation (MEE) testing may be conducted by the radiologic technologist under medical physicist oversight. Instead, FFDM manufacturers are required to notify FDA of their new software upgrades for which the MEE may be conducted under medical physicist oversight. FDA will then simply acknowledge receiving this notification, and the manufacturer may then proceed with disseminating the upgrade. 1. GE Healthcare - Effective 10/29/07Several software upgrades identified as:
The first four of these upgrades are intended to address image presentation issues on various versions of the SenoAdvantage workstation. The fifth is an upgrade of the software used to calibrate the displays. The issues addressed by these upgrades are described in more detail in the MEE instructions to be provided to the facility. 2. Hologic - Effective 11/21/07Software upgrade v. 3.3 as described in “Selenia Release Notes 3.3.1.1” - The release notes include the description of the upgrade, the model, and all tests that are required to evaluate the equipment after the software change (upgrade). The Release Notes are provided to mammography facilities to inform them of which tests are to be performed for the MEE after installation, and as information regarding the Selenia software change. They include revisions back to the “Selenia 3.1.2” version and they are not meant to replace the Selenia operator’s manual. Furthermore, this software upgrade allows the configuration of the Selenia FFDM system with a tungsten (W) x-ray tube and rhodium (Rh) and silver (Ag) filtration. 3. GE Healthcare - Effective 3/14/2008A software upgrade identified as:
This software upgrade applies to the Seno Advantage 1.0, 1.1, and 1.2 Workstations that are intended for final interpretation of clinical mammograms. The issues addressed by this upgrade are described in more detail in the MEE instructions that will be provided to the facility. 4. Fuji Systems - Effective 4/2/2008A retroactive software upgrade identified as:
This software upgrade does not require any testing by the facility prior to implementation and therefore does not require MEE testing under medical physicist oversight as called for under the amended Alternative Standard # 6. Note: This upgrade was reviewed and approved by FDA's Office of Device Evaluation (ODE) in May 2007 as part of a pre-market approval (PMA) device modification submission. 5. Fuji Systems - Effective 4/25/08A retroactive software upgrade identified as: "Fuji IIPm (CR-IR348) Version 5.0(B) Software Changes." This may also be referred to as a software upgrade V5 for FCRm. This upgrade applies to the FCR Image Console (lIP), which is a dual modality workstation used for both FCRm digital mammography and general radiographic (CR) studies. It includes new functions, enhancements, and corrections in both areas. The following items of the upgrade are intended for "performance correction" and are the only ones that apply to mammography in the United States:
The Fuji representative, Customer Solutions Engineer (CSE), has/will perform a reboot of your PC from the start menu and verify that the most current software version number is indicated on the initialization screen. This upgrade does not require any testing by the facility prior to implementation; therefore, it does not require mammography equipment evaluation (MEE) testing under medical physicist oversight as called for under amended Alternative Standard #6. 6. General Electric Seno Advantage Diagnostic Workstations – Effective 6/20/08These upgrades will be distributed as Field Modification Instruction (FMI) 12092, “PV3 for Seno Advantage 2.0 and 2.1.” The software upgrades address image presentation issues on various versions of the Seno Advantage workstation. The issues addressed by these upgrades are described in detail in the Mammography Equipment Evaluation (MEE) instructions to be provided to the facility. They are identified by General Electric (GE) Healthcare as SA2_08.1 and SA21_05.1 The software upgrades are intended to improve the operation of image processing algorithms for some exceptional cases and there are no routine quality control (QC) tests intended to directly evaluate the system performance under such cases. A checklist will be provided by the installer informing the operator which upgrade was performed and what MEE tests need to be done. GE Healthcare believes that all of the specified tests can be performed by the QC Technologist. None of the tests are required to be done by a medical physicist. The facility’s post-installation responsibilities are included in the information to be provided to each facility. The software upgrade identified as SA21_05.1 affects the content of the QC manual for the Seno Advantage workstation. This upgrade facilitates the operation of the workstation with either Siemens monitors or Barco monitors, and the QC manual has been revised to accommodate the differences in the details of the QC software between these two monitor types. No tests have been added or deleted. No action limits have been changed. A copy of the new revision of the QC manual will be included with upgrade SA21_05.1. While it will be “backward compatible” and could be used by facilities that do not have Barco displays installed, there is no need for facilities with Siemens monitors to obtain a copy of the later revision. 7. General Electric Senographe 2000 D FFDM Systems – Effective 7/14/08This upgrade is identified as Field Modification Instruction (FMI) 12063A, “Senographe 2000 D AWS v17.5 Upgrade.” This upgrade will:
It is anticipated that when this upgrade is performed, in over 90% of the systems, no detector calibration information will be lost and that there will be no change in patient dose. However, in some cases it will be necessary to upgrade the Image Detection Controller (IDC). Following this upgrade the Detector Conversion Factor and AOP calibrations, among others, must be done, and these calibrations could affect patient dose. For those cases where this occurs, the facility will be informed by the installer that the Alternative Standard may not be applied and the facility’s physicist must be contacted to perform an in-person Mammography Equipment Evaluation (MEE). By means of a checklist, the facility will be informed of the extent of the upgrades and the effect of the upgrades on the facility’s responsibilities regarding the MEE. Following this software upgrade, the user will be instructed to check that the label regarding view and laterality labeling has been affixed to the radiation shield of the mammography system, and to print films from the AWS that demonstrate correct labeling of the images. The user is further instructed to acquire a phantom image and score it on all display devices. 7. Combined Mammography Medical Outcomes Audit for Multiple Mobile Mammography UnitsFDA approved this alternative standard on November 4, 2002 and amended it on December 3, 2002. The amended alternative standard became effective on the latter date. It has no time limit. Some accreditation bodies accredit each mobile unit separately, even if two or more units are under the same ownership. This approach leads to each mobile unit being certified as a separate facility. Therefore, before the alternative requirement became effective, a separate mobile medical outcomes audit had to be performed for each unit. This alternative allows owners of multiple mobile mammography units to perform a combined mammography medical outcomes audit for all of the units if the specified conditions are met. The original standard is 21 CFR 900.12(f)(1), which states:
The approved alternative, as amended, is:
8. Separate Assessment of Findings For Each BreastFDA approved this alternative standard on July 3, 2003. It became effective on that date. It has no time limit. It allows the interpreting physician to provide a separate assessment of findings for each breast in the medical report, without the need to also provide an overall assessment of findings. Therefore, the interpreting physician can choose between providing separate assessments under this alternative or providing an overall assessment for the examination under the original standard. The original standard is 21 CFR 900.12(c)(1)(iv), which states:
The approved alternative is:
The conditions of use of this alternative are that:
9. Correction Period When Components of the Selenia Full Field Digital Mammography System Fail Quality Control TestsThis alternative requirement was approved on August 21, 2003 and became effective on that date. It was amended on October 10, 2007. It has no time limit. It allows a 30 day period for corrective actions following the failure of specified quality control tests by the Selenia Full Field Digital Mammography System. The specified tests are equivalent to quality control tests for screen-film systems for which a 30 day correction period is already allowed. The alternative standard also divides into two groups the quality control tests whose failure requires corrective action before the failing component is used again during patient examinations. This division makes it clear that when the test failure is related to the acquisition of images only, image acquisition must cease until the problem is corrected but image interpretation can continue. Similarly if the test failure is related to the interpretation of images, image acquisition can continue but image interpretation with the failed component must cease until the problem is corrected. The original standard is 21 CFR 900.12(e)(8)(ii), which states:
The approved alternative is:
10. Amendment to the Alternative Requirement for the Correction Period When Components of the Senographe™ 2000D Full Field Digital Mammography (FFDM) System Fail Quality Control TestsThe following amended alternative standard was approved on August 31, 2007 and became effective on the date of approval. It has no time limit. It supersedes Amended Alternative Requirement # 10 which was approved August 25, 2003. The current amendment adds the Sub-System MTF Measurement test to the quality control tests already listed in the 2003 amendment. The 2003 amendment replaced the specific reference to the GE Senographe™ 2000D FFDM system with a generic reference to an "FDA-approved GE" FFDM system. Like the original standard, it allows a 30 day period for corrective actions following the failure of specified quality control tests by an FDA-approved GE FFDM system. However, it divides into two groups the tests whose failure requires corrective action before the failing component is used again during patient examinations. This division makes it clear that when the test failure is related to the acquisition of images only, the review of already acquired images can continue and when the test failure is related to the image review components only, images can continue to be acquired. In approving the amendment, FDA stated that if GE introduces new FFDM systems having QC tests other than what is included in the original or amended standard, the amended alternative standard would not be applicable to such systems. The original standard is 21 CFR 900.12(e)(8)(ii), which states:
The approved alternative standard is an alternative is:
11. Modifications in the Assessment Categories Used in Medical ReportsFDA approved two alternative requirements dealing with the content of the medical report on August 29, 2003. They became effective on that date. They do not have a time limit. One of these adds a new assessment category for use in the reports of the mammography examinations and also adds clarifying language to the existing assessment categories. The second adds a reference to the possible need to obtain prior mammograms in order to make a final assessment. The original standards are 21 CFR 900.12(c)(1)(iv) and (v), which state:
The approved alternatives are:
As was the case with the original standard, only the words in quotation marks are required to be included in the medical report when giving the assessment category or indicating that no final category can be assigned at the present time. The remaining language is intended to provide explanations of the categories in order to promote their consistent use. It is not required to be included in the medical report, although the interpreting physician may do so if he or she wishes. 12. Assessment category for "Post Procedure Mammograms for Marker Placement"This alternative standard was approved and became effective on September 17, 2003. It has no time limit. This approved assessment category can only be used for a post procedure mammogram to confirm the deployment and position of a breast tissue marker. The lay summary, which must be provided to the patient, must be specific to the procedure. If the facility makes this post procedure examination part of the interventional study instead of a separately charged examination, then it does not fall under MQSA and the approved alternative standard does not apply. The original standard is 21 CFR 900.12(c)(1)(iv), which states:
The granted alternative is:
(v) In cases where no final assessment category can be assigned due to incomplete work-up, "Incomplete: Need additional imaging evaluation" shall be assigned as an assessment and reasons why no assessment can be made shall be stated by the interpreting physician; and... 13. System Artifact Testing at Remote Mobile Mammography Sites Where Film Processing Takes Place Using Processors Permanently Located at the Site.This alternative standard was approved and became effective on December 14, 2004. It has no time limit. It covers the system artifact testing at remote sites where mobile mammography facilities utilize processors that are permanently located at those sites. The approved alternative permits a specially trained quality control technologist to make system artifact films and phantom images at remote processor sites used by mobile mammography facilities and submit them to the facility medical physicist for evaluation. This relieves the facility from the need to have the medical physicist visit each remote processing location as a part of the annual survey. Although this request was approved for a specific mobile facilities operator, it has been extended to cover any other mobile facility that elects to apply it and meets the conditions described below. The original standard states:
The above is required to be performed annually by a qualified mammography medical physicist under the following:
The granted alternative is:
Conditions of this alternative requirement are: The medical physicist will train the mammography technologist in the procedures as outlined in the American College of Radiology (ACR) Mammography Quality Control Manual, 1999. The training will include the direct supervision of that mammography technologist while she carries out these procedures at a mobile mammography site. The medical physicist will also carry out the identical procedures at that site. These procedures are:
The medical physicist will evaluate and compare all these test films produced by the mammography technologist and by the medical physicist. If the results are comparable and the medical physicist reaches the same conclusions after reading the films produced by the mammography technologist that would be reached reading the physicist’s films, then the mammography technologist is considered to be capable of producing adequate films for this test without direct supervision. At least once a year, the physicist will perform all required annual survey items including the phantom image and artifact test at one location serviced by the medical physicist and will use the technologist’s ACR phantom and absorber materials to perform the on-site testing.Prior to this annual visit, the mammography technologist will take two artifact films using four cm of absorber provided by the physicist at each remote processing site. The films will be developed according to the procedures in the 1999 ACR manual. The artifact test procedures will be repeated for each target/filter combination and each processor used clinically at each site. The mammography technologist will also take a phantom image at each site using the facility’s ACR phantom. The images taken by the mammography technologist at the remote sites will be brought to the chosen annual survey site for evaluation by the medical physicist during the annual survey. Applicable records for the physicist’s “Evaluation of Site’s Technologist’s QC Program” as described in the 1999 ACR Mammography Quality Control Manual, will also be provided from each remote site for the physicist’s review. 14. System Artifact Testing of Target Filter CombinationsThis alternative standard was approved and became effective on June 16, 2005. It has no time limit. It covers the system artifact testing various target filter combinations. The approved alternative permits the system artifact test to be performed without testing all target filter combinations during the annual physics survey. The alternative standard covers any facility that elects to apply it and meets the conditions described below. The original standard states: The granted alternative is: Conditions of this alternative requirement are:
|
FDA approved and made effective this alternative requirement on August 2, 2006. It has no time limit. It allows a 30 day period for corrective actions following the failure of specified quality control tests by the FCRm Full Field Digital Mammography System. The specified tests are equivalent to quality control tests for screen-film systems for which a 30 day correction period is already allowed. The alternative standard also divides into two groups the quality control tests whose failure requires corrective action before the failing component is used again during patient examinations. This division makes it clear that when the test failure is related to the acquisition of images only, image acquisition must cease until the problem is corrected but image interpretation can continue. Similarly if the test failure is related to the interpretation of images, image acquisition can continue but image interpretation with the failed component must cease until the problem is corrected.
The original standard is 21 CFR 900.12(e)(8)(ii), which states:
21 CFR 900.12(e)(8): Use of test results
(ii) If the test results fall outside of the action limits, the source of the problem shall be identified and corrective actions shall be taken:
(A) Before any further examinations are performed or any films are processed using the component of the mammography system that failed any of the tests,
described in paragraphs (e)(1), (e)(2), (e)(4)(i), (e)(4)(iii), (e)(5)(vi), (e)(6), or (e)(7) of this section;
The approved alternative is:
21 CFR 900.12(e)(8): Use of test results
(ii) If the test results for the FCRm FFDM system fall outside the action limits, the source of the problem shall be identified and corrective actions shall be taken:
(A) If any of the following quality control tests that evaluate the performance of the image acquisition components of the FCRm FFDM system produces results that fall outside the action limits as specified by the manufacturer, the source of the problem shall be identified and corrective action shall be taken on the component(s) that caused the failure before any further examinations are performed using the failed component(s). If the component(s) that caused the failure (e.g. exposure unit, CR reader, display device) is replaced by an alternative device and the mammography system passes the re-test, image acquisition and interpretation may continue using that combination of devices.
(1) CNR Weekly Check
(2) Phantom Image
(3) Compression
(4) System Resolution
(5) CR Reader Scanner Performance
(6) Imaging Plate (IP) Fog
(7) Image Quality Evaluation
(8) Dynamic Range
(9) Primary Erasure
(10) Dose
(11) Post-move and pre-examination tests (CNR Weekly Check and Phantom Image) for mobile FCRm FFDM systems(B) If any of the following quality control tests that evaluate the performance of a diagnostic device used for mammographic image interpretation (i.e. laser printer, physician's review station) produces results that fall outside the action limits as specified by the manufacturer, the source of the problem shall be identified and corrective action shall be taken before that device can be used for mammographic image interpretation. Clinical imaging may be continued. If available, an alternative approved diagnostic device may be used for mammographic image interpretation:
(1) Phantom Image
(2) Monitor QC
(3) Printer QC
(4) CR Reader Scanner Performance
(5) Image Quality Evaluation
(6) Dynamic Range
(7) Viewing and Viewing Conditions(C) If any of the following quality control tests that evaluate the performance of other X-ray room or exposure function or the diagnostic devices used for mammographic image interpretation produces results that fall outside the action limits as specified by the manufacturer, the source of the problem shall be identified and corrective action shall be taken within thirty days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period:
(1) Visual Checklist
(2) Repeat Analysis
(3) Mammographic Unit Assembly Evaluation
(4) Collimation Assessment
(5) Automatic Exposure Control (AEC) System Performance Assessment
(6) System Artifact Evaluation
(7) kVp Accuracy and Reproducibility
(8) Beam Quality Assessment and Half –value Layer Measurement
(9) Radiation Output
(10) Inter-Plate Consistency
(11) S value Confirmation
Updated July 30, 2008
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH