Engineering Diffractometer - Equipment
|
VULCAN Diffractometer Schematic
Click image for a larger view.
|
At present, this instrument is tentatively named VULCAN, after
the Roman god of fire and metalworking. Although the primary
use of VULCAN is intended for deformation and residual stress
related studies, other uses include spatial mapping of chemistry,
microstructure, and texture.
The desired performance for VULCAN as determined by the user
community is listed below.
- rapid volumetric (3-dimensional) mapping with a sampling
volume of 1 mm3 and a measurement time of minutes
- very high spatial resolution (0.1 mm) in one direction with
a measurement time of minutes
- ~20 well defined reflections for in-situ loading studies
- ability to study kinetic behaviors in sub-second times
- simultaneous characterization capabilities, including dilatometry,
weight, and microstructure
- ancillary equipment such as a furnace and load frame be an
integrated part of the instrument
Together, these requirements call for a "compound" engineering diffractometer
with a large degree of flexibility for intensity-resolution optimization.
The design philosophy is therefore to deliver a diffractometer with
the highest desirable Q-resolution over a large angular range. For
experiments that do not require such a high resolution, the incident
beam divergence can be relaxed for intensity gain at the sample position.
The key to this design is an interchangeable guide-collimator system.
This is a 44-m long instrument, consisting of a 3 m in-monolith guide,
a frame definition chopper, a 20 m curved guide followed by a 12 m
straight guide, and a 5 m interchangeable guide-collimator system.
All guides are m=3 supermirror guides and have a cross-section of 12×50
mm2.The interchangeable guide-collimator system consists
of five-1 m segments, each containing two channels which is either
a m=3 supermirror guide or a straight collimator. Each channel can
be translated into the beam position at the push of a button. These
channels are used in combinations to produce an incident beam of desirable
divergence. A 1 m space is allowed between the sample and the exit
of the guide-collimator system. In addition to housing the incident
slit unit, this space may also be used to accommodate additional shielding
or collimators. The detectors cover 60° - 150° in 2q and ±30° in
the vertical plane. They are located on a locus designed to give an
approximately equal resolution across the covered 2q range.
The radius of the detector locus ranges from 1.5 m at 150° to 5
m at 60°.
VULCAN Diffractometer at SNS
Click image for a larger view.
|
|
VULCAN can be operated flexibly for experiments having very different
intensity-resolution requirements. In the high intensity mode, the
m=3 supermirror channel will be employed in all segments of the guide-collimator
system, effectively extending the supermirror guide all the way to
1 m from the sample. This is the preferred mode of operation for most
stress mapping experiments where the specimens already exhibit significant
peak broadening due to deformation, for kinetic studies where time
resolution is critical, and for texture studies where a high Q-resolution
is unnecessary. Note that even in this mode, the resolution is still
quite good, ranging from 0.2% for 2q=150º to
0.65% for 2q=60° at d=1 Å.
For comparison, the resolution for ENGIN, currently the only dedicated
engineering diffractometer at a pulsed neutron source, is 0.7%. For
in-situ loading studies where one needs to examine the response of
individual reflections, however, it may be necessary to operate the
diffractometer in the high resolution mode. In this mode, the collimator
channel will be employed in all segments of the guide-collimator system,
which produces a resolution of 0.10-0.22% for all detectors.
This resolution is already better than that of the 148° (highest
resolution) detector bank on GPPD at IPNS, which is sufficient to resolve
~ 20 peaks for cubic materials, such as Fe and Al, and for low-symmetry
materials, such as Al2O3. Table I summarizes
the performance matrix for these two vastly different operation modes.
If the spatial variation is not of interest, as is the case for most
in-situ loading experiments, the loss of neutron intensity can be more
than compensated with the use of a large sample. For a 60 Hz target,
the wavelength bandwidth is Dl~1.3 Å,
which can be positioned anywhere in the incident beam spectrum. If
necessary, this bandwidth can be doubled by eliminating every other
pulse using the frame definition chopper, effectively operating the
instrument at 30 Hz.
Table I Performance matrix for
selected instrument configurations for a peak at d=1Å. |
|
d=1Å |
High Intensity
Mode (*) |
High Resolution
Mode |
Detector 2q(º) |
l (A) |
Relative Flux (%) |
Dd/d
(%) |
Relative Flux (%) |
Dd/d
(%) |
60 |
1.000 |
100 |
0.55 |
31 |
0.21 |
90 |
1.414 |
100 |
0.38 |
27 |
0.20 |
120 |
1.778 |
100 |
0.27 |
25 |
0.18 |
150 |
1.932 |
100 |
0.18 |
23 |
0.15 |
SANS Q range |
|
0.01-0.18 Å-1 |
|
(*) Base line reference
for intensity comparison. |
A 20×20 mm2 two-dimensional position sensitive detector
will be mounted 4 m downstream of the sample for simultaneous diffraction
and small angle measurements. By using a small specimen (e.g., < 5
mm width), a poor-man’s SANS can be realized with little disruption
to the base-line design. The Q-range was estimated to be 0.01 Å-1-0.18 Å-1.
When desirable, a larger Q-range can be obtained by eliminating every
other pulse using the frame definition chopper. A separate detector
can be mounted at the center of the SANS detector to record transmission
data. By analyzing Bragg edges in the transmission data, it is possible
to obtain information about the chemistry and lattice strains in materials
in the incident beam path. While methods for systematic analysis of
the transmission data have not been fully developed, this technique
has the potential for making measurements an order of magnitude faster
than with the conventional diffraction method.
A great challenge lies in developing the technology that enables
diffraction measurements with a 0.1 mm spatial resolution. Radial collimators,
currently the method of choice at pulsed neutron sources for defining
the sampling volume, become inefficient when used to define a 0.1 mm
wide diffracted beam. This is because with today’s technology,
the minimum blade thickness is 0.1 mm. As part of the research efforts,
two technologies will be pursued in parallel to achieve the 0.1 mm
spatial resolution, one based on reducing the blade thickness of radial
collimator and the other based on a novel Bragg mirror imaging concept.
Xun-Li Wang is the SNS instrument scientist responsible for the design and construction
of the Engineering Diffractometer. To find out more about him, visit his webpage.
While the primary use of this instrument is intended for deformation and residual
stress related studies, other uses might include spatial mapping of chemistry,
microstructure, and texture. Furnace, load frame, and other auxiliary equipment
for in-situ and time-resolved measurements will be an integrated part of the
instrument. A workshop was held on January 20-21, 2000 in Atlanta to define
the required performance matrix.
The following documents contain more detailed information about the
expected performance and design of the diffractometer:
Science Case For
the Vulcan Diffractometer (PDF 99.1KB)
Summary of the
Breakout Session on Basic Mechanical Properties (PDF 200KB)
Conceptual Design
Report (PDF 201KB)
IDT Meeting Report
(March 7, 2001)
IDT Meeting Report
#1 (Nov. 18-19, 2002), Oak Ridge — T. M. Holden
Instrument data sheets
(PDF)
The following personnel are currently working on the diffractometer:
rennichgq@sns.gov, Engineer
turnerb@sns.gov, Designer
|