This guidance was written prior to the February 27, 1997 implementation of FDA’s Good Guidance Practices, GGP’s. It does not create or confer rights for or on any person and does not operate to bind FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute, regulations, or both. This guidance will be updated in the next revision to include the standard elements of GGP’s.

Skip NavigationFDA Logo links to FDA home pageCenter for Devices and Radiological Health, U.S. Food and Drug AdministrationHHS Logo links to Department of Health and Human Services website

FDA Home Page | CDRH Home Page | Search | CDRH A-Z Index | Contact CDRH U.S. Food and Drug Administration Center for Devices and Radiological Health
horizonal rule

GUIDANCE ON THE CONTENT OF PREMARKET NOTIFICATION [510(K)] SUBMISSIONS FOR EXTERNAL INFUSION PUMPS


Related Information

 


                 
                                March 1993

        GUIDANCE ON THE CONTENT OF PREMARKET NOTIFICATION [510(K)]
                 SUBMISSIONS FOR EXTERNAL INFUSION PUMPS

I.   Introductory Information

A.   Scope

     This document establishes the 510(k) review requirements for external 
infusion pumps.  The external infusion pump may be electrically powered, 
mechanically powered, or use an alternate energy source.  Examples of 
devices within this generic type include peristaltic, piston, 
spring-driven, syringe, and elastomeric types of pumps.  Accessory devices, 
such as carrying cases, are also included within this type of device.  
Disposable functional components such as infusion sets or cassettes, 
provided or specifically recommended for use with these devices, must be 
identified and demonstrated to be either:  1) substantially equivalent, or 
2) legally marketed medical devices.  If these functional components are 
included for evaluation in this submission they should meet all other 
applicable guidance.  

     EXCLUSIONS

     This document does not address submissions for implantable infusion 
pumps or closed-loop devices, which both require premarket approval.  This 
document also does not cover pressure infusors for I.V. bags (as described 
in 21 CFR 880.5420).

B.   Purpose

     This guidance is intended to:

     1.   assist persons (manufacturers, distributors, or
          importers) in organizing premarket notifications for
          external infusion pumps; 

     2.   achieve consistency in meeting of requirements and in
          the presentation of information; and

     3.   guide FDA review staff in conducting and documenting
          the review of premarket notifications for external
          infusion pumps.

C.   Definitions

     1.   Infusion Pump:  described in FDA regulation, 21 CFR
          880.5725, as "a device used in a health care facility
          to pump fluids into a patient in a controlled manner. 
          The device may use a piston pump, a roller pump, or a
          peristaltic pump and may be powered electrically or
          mechanically.  The device may also operate using a
          constant force to propel fluid through a narrow tube
          which determines the flow rate.  The device may include
          means to detect a fault condition, such as air in, or
          blockage of, the infusion line and to activate an
          alarm." 

     2.   Cassette:  a specially designed disposable
          component/accessory that mates with the infusion pump
          to complete the pumping or controlling mechanism. 

     3.   Elastomeric Pump:  an infusion pump which utilizes the
          energy in an elastic membrane to provide the force for
          fluid delivery.

     4.   Flow Profile:  a graph or table representing
          numerically the flow rate versus time over the course
          of an infusion.

     5.   Fluid Pathway:  all portions of the device or any
          accessories which contact the fluid being infused.

     6.   Patient Controlled Analgesia (PCA) Pump:  an infusion
          pump, intended for the delivery of analgesics, which is
          equipped with a component which allows for additional
          limited delivery upon patient demand. 
  
     7.   Syringe Pump:  an external infusion pump which utilizes
          a piston syringe as the fluid reservoir and to control
          fluid delivery.     

     8.   Intended Use:  the objective intent of the persons
          legally responsible for the labeling of the device. 
          The intent is determined by their expressions or may be
          shown by the circumstances surrounding the distribution
          of the device.  The objective intent may, for example,
          be shown by labeling claims, advertising matter, or
          oral or written statements by such representatives.  It
          may be shown by the offering or the using of the
          device, with the knowledge of such persons or their
          representatives, for a purpose for which it is neither
          labeled nor advertised ( 801.4).  Some use conditions
          for infusion pumps may include single use only, home
          use, reusable, intravenous and intra-arterial delivery.

     9.   Abbreviations:
          AAMI - Association for the Advancement of Medical
                Instrumentation
          ANSI - American National Standards Institute
          CBER - Center for Biologics Evaluation and Research
          CDER - Center for Drug Evaluation and Research
          CDRH - Center for Devices and Radiological Health
          CFR - Code of Federal Regulations
          DSMA - Division of Small Manufacturers Assistance
          FDA - Food and Drug Administration
          FR - Federal Register
          IEC - International Electrotechnical Commission
          ISO - International Organization for Standardization
          OCS - Office of Compliance and Surveillance
          ODE - Office of Device Evaluation
          SMDA - Safe Medical Devices Act of 1990
          UL - Underwriters Laboratory

D.   General Principles Regarding Presentation of Data

     1.   Editorial Considerations:  The 510(k) should be
          carefully edited, as well as scientifically reviewed
          before it is submitted to FDA.  It should be proofread
          to assure that all pages/sections are included and are
          properly indicated, consecutive, distinctly copied, and
          legible.

     2.   Abbreviations:  Standard abbreviations acceptable to a
          significant peer reviewed journal should be used
          wherever possible.  All other abbreviations should be
          identified at the beginning of each section in which
          they are used or in footnotes to tables and graphs.

     3.   Data Availability:  This document outlines typical
          circumstances of data review.  It is not possible to
          anticipate all situations that may require FDA review. 
          Thus, those submitting applications should be aware
          that they may be asked to submit additional data, to
          present data in another format or to provide more
          detailed explanations of the information submitted, if
          required to establish equivalence.

          Applicants should keep data used for the 510(k)
          submission on file in a controlled and well-organized
          format.  This will allow the applicant to expeditiously
          supply FDA with additional information or analysis if
          required.  Errors in data that are identified by the
          applicant after submission to FDA should be brought to
          FDA's attention immediately.

     4.   Tables and Graphs:  Well-constructed tables are
          fundamental to the reporting and evaluation of data. 
          All tables should be clearly identified and captioned
          with symbols keyed to a footnote or accessible
          reference page that adequately indicates the nature of
          the data.

          Graphs should supplement, not replace, data tables. 
          They should be of a high quality.
        
     5.   Published Literature:  Published methods or data
          referenced in study reports should be appended to the
          study report.  Reprints of other referenced published
          reports or data should be appended to the section in
          which they are referenced.  All referenced reports and
          data should be summarized including an explanation how
          it relates to the current submission.  Reference
          citations should be complete (e.g., title, author,
          volume, year).

     6.   Protocols and Data Analysis:  Test reports must include
          the protocol (objectives, precise description of
          materials, experimental methods, controls),
          observations, statistical methods and analyses,
          conclusions and comments.  Do not submit raw data. 
          Additional specific directions on protocols are
          included in sections that follow.
          
     7.   Reference to Submitted Data:  In support of the 510(k)
          the applicant may reference any information previously
          submitted to FDA.  If the applicant did not submit the
          referenced data he must provide, or have the submitter
          provide to FDA, a letter of authorization.  Often, if
          the data are not extensive, resubmitting data in the
          510(k) will facilitate the review of the document.

E.   Document Availability

     The following documents are available from DSMA [(800)638-2041 or 
(301)443-6597]:

       Tripartite Biocompatibility Guidance for Medical Devices   ODE Blue 
Book Memorandum #K90-1:  510(k) Sterility Review Guidance   Reviewer 
Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review.

II.  Content and Organization of Information in a 510(k) for an
     External Infusion Pump
   
A.   Cover Letter

     The submission shall have a signed cover letter providing the 
following information described in  807.87 (Information required in a 
premarket notification submission):
    
     1.   The infusion pump's trade or proprietary name.
                              
     2.   Common Name:  Syringe Pump, Elastomeric Pump, etc.
            
     3.   Classification name:  Infusion Pump

     4.   The establishment registration number, if applicable,
          of the sponsor, owner or operator submitting the
          premarket notification
               
     5.   Class: II
          Panel: 80
          Procodes:  list all the following that apply
                     FRN - Infusion Pump
                     MEA - PCA External Infusion Pump
                     MEB - Elastomeric External Infusion Pump
                     LZG - External Insulin Infusion Pump
                     LZH - External Enteral Infusion Pump   
                     FPA - Intravascular (IV) Administration Set

     6.   A statement explaining the purpose of the submission
          (e.g., new device, significant modification of device
          previously found equivalent (new intended use,
          material, or manufacturing process, etc.)).  Refer to
           807.87(g) for additional information regarding changes
          to devices.  The change may require some or all of the
          information needed for a new device.  Please supply the
          previous 510(k) number(s), if applicable.

     7.   A brief statement indicating the device is similar to
          and/or different from other products of comparable type
          in commercial distribution.

     8.   Name, address, and phone number of a U.S. contact
          person, if available.

B.   Labels and Labeling

     1.   The submission shall contain proposed labels, labeling,
          and advertisements sufficient to describe the device,
          its intended use, and the directions for use.  Labels
          include the information affixed directly to the device
          or its container or packaging.  Labeling also includes
          professional or patient package inserts, and any other
          information that accompanies the device.  

     2.   The labeling must meet the requirements of 21 CFR Part
          801 as it relates to a determination of intended use. 
          ODE will concentrate on the following portion of Part
          801:

               Subpart A,   801.4 and 801.5, related to intended
               uses and adequate directions for use; and

               Subpart B,   801.109 and 801.116, related to
               prescription devices and commonly known
               directions.  

          Other provisions of Part 801 are deferred for review to
          CDRH/OCS Device Labeling Compliance Branch.
     
     3.   Labeling for an infusion pump should include: 

          a.   a list of device specifications including, but not
               limited to:  reservoir volume, flow rates and
               profiles (if applicable), accuracy, residual
               volume, and the operational conditions (e.g.,
               temperature, pressure, fluid viscosity) under
               which these are valid; 

          b.   route(s) of administration as indicated in the
               statement of intended use (see section D.2.);

          c.   the fluid(s) to be administered by the pump as
               indicated in the statement of intended use (see
               section D.2.);

          d.   comprehensive directions for preparation and use
               for all possible functions of the device;

          e.   a description of all warning and alarm features;

          f.   a copy of patient instructions for use if the
               device is intended for home use, or would
               reasonably be expected to be used in a home
               setting; 
          
          g.   the prescription statement found under
                801.109(b)(1);  
     
          h.   instructions for cleaning and maintenance which is
               to be done by the user; and 

          i.   an identification of any dedicated administration
               set or the specifications and/or specific models
               of infusion sets which are appropriate for use
               with this pump.

C.   Standards

     Listed are some, but not all, of the standards related to infusion 
pumps:

     1.   AAMI, Draft Infusion Device Standard

     2.   UL 544 Standards for safety, medical and dental
          equipment

     3.   IEC 601-1/ANSI ES1-1985 Safe Current Limits for
          Electromedical Apparatus

     The applicant may certify that the device meets a standard. The 
applicant then is obliged to comply with the standard and maintain 
documentation of tests showing that the device meets the standard.  
Certification of meeting a specific standard and reference to standards in 
the 510(k) may reduce the documentation needed in the 510(k) submission.  
This is noted in pertinent sections.   

D.   Device Description
     
     The applicant must submit a complete description of the device, 
including all models and variations.  

     1.   State the type of infusion pump (e.g., peristaltic,
          piston, syringe).  Provide a labeled representation of
          the device in sufficient detail to facilitate the
          evaluation of the nature and operation of the device
          (e.g., photographs, detailed drawings, or engineering
          drawings).  If the labeling already includes sufficient
          illustrations of the device, refer to the labeling.
     
     2.   Provide a clear statement of the intended use(s) of the
          infusion pump. 
          
          a.   Specify the route(s) of administration.

          b.   If the infusion pump is labeled for use with a
               specific drug/biologic, the applicant must supply
               information demonstrating that use of the
               drug/biologic with the device is consistent with
               the approved drug/biologic labeling.

          c.   Identify any specific uses for the pump.  For
               example, PCA is a generally accepted specific use
               for a pump used to deliver analgesics.

          d.   State whether the infusion pump is indicated for
               the delivery of blood or blood products.

     3.   Provide the following information for the device.  The
          applicant may refer to relevant standards.

          a.   Physical Specifications and Descriptions

          (1)  Components:  Identify and describe all major
               components or component groups (i.e., not
               individual electronic or mechanical components)
               and give dimensions.  Each component/component
               group should be identified as reusable or
               disposable.  Identify any component/component
               group which contacts the fluid path.

          (2)  Pumping Mechanism (Fluid Control Mechanism): 
               Describe the physical method used to generate and
               control fluid flow.  Identify and provide
               specifications for all major components involved
               in this process.

          (3)  Administration Sets & Drug Reservoir:  Identify
               sets appropriate for use with the pump. 
               Peristaltic and piston devices often have
               specialized administration sets which must be used
               with the individual pump.  Describe the drug
               reservoir.  Indicate if these are currently
               legally marketed devices or if they are included
               for evaluation in this submission.   

          (4)  Power Requirements:  Describe the energy/power
               requirements and characteristics.  Indicate if
               device meets UL 544 (if applicable).  If not,
               describe how it deviates from the standard.  

          (5)  Materials:  Identify the basic materials of
               construction.  Provide the chemical formulation of
               all materials which contact the fluid path.  

          b.   Operational Specifications and Descriptions

          (1)  Flow Rates and Profiles:  State the accuracy for
               the device.  Describe all possible delivery rates
               and flow profiles, and the operational conditions
               (e.g., temperature, pressure, fluid viscosity)
               under which the device will meet the stated
               accuracy figure.  List the reservoir volume and
               residual volume.

          (2)  Safety/Alarm Functions:  For electronic pumps
               describe all of the various alarm, warning and
               safety functions, both hardware and software, of
               the device.  For mechanical devices describe any
               safety mechanisms built into the device.

          (3)  Other Capabilities:  Describe any other
               capabilities the device may possess.

          c.   Biological Specifications

               State the biocompatibility category for all
               components and materials in the fluid pathway per
               the draft ISO 194 Biocompatibility Standard or the
               Tripartite Biocompatibility Guidance.

          d.   Chemical Specifications 

          (1)  State the compatibility requirements for the
               device with any specific drug or biologic
               referenced in the device labeling.

          (2)  State the stability requirements for storage of
               any specific drug or biologic referenced in the
               device labeling.

E.   Descriptive Comparison to a Legally Marketed Device

     Identify a legally marketed infusion pump to which substantial 
equivalence is claimed.  (If possible, identify the 510(k) number(s).)  
More than one infusion pump can be listed, but the device(s) chosen should 
be as close in intended use and technology to the new device as possible. 
Provide the information noted below to show how the new device is both 
similar to or different from the legally marketed device.  Side by side 
comparisons, whenever possible are desirable (see Attachment 2).  This 
information may be identical to that provided under Part D and the 
applicant may wish to combine some or all of Parts D and E information.  
Indicate how the differences may affect safety and effectiveness.    

     1.   Provide labeling (labels, instructions for use,
          promotional material) for the legally marketed
          device(s) to which substantial equivalence is claimed. 
          To facilitate comparison, also include clear
          representations of the legally marketed device(s),
          unless the labeling has ample information.
     
     2.   Compare and contrast the intended use for the new
          device to the predicate device(s).  

     3.   Compare materials used to fabricate fluid path
          components.  The precise materials of the fluid path of
          the new device, and if possible, the predicate
          device(s) should be identified to the extent possible.

     4.   Compare the operational principles including mode of
          action. 

     5.   Compare physical, operational, biological, and chemical
          specifications.

F.   Performance Data Supporting Substantial Equivalence

     Provide the protocols and results of the tests indicated below.  If 
the stated test is taken from a standard that specifically addresses the 
performance criterion, then the applicant should reference the standard and 
certify that the device will meet the criterion.  Data need not be 
submitted in this instance.
  
     The studies should be well-designed to meet the stated objectives.  
This will include rigorous attention to: statistical elements (hypotheses, 
test statistics, analyses, sample size and sampling, power, etc.), 
inclusion/exclusion criteria, controls, minimization of bias, test 
parameters (endpoints), follow-up, evaluation criteria, etc.  Some of the 
above points may overlap.  Ample reference material exists on study design 
and methods upon which the applicant may rely (e.g., biocompatibility). 
          
     1.   Bench Data

          The following bench data is generally necessary for a
          non-electronic pump or any pump with a unique design
          unless the sponsor can completely  address all of the
          following issues using device specifications. 
               
          a.   For a pump intended only to maintain a constant
               set flow rate, this data should demonstrate that
               the device can maintain a set flow rate over the
               complete course of the infusion within the
               designated accuracy.  The testing should
               demonstrate adherence to specifications at the
               limits of the operational parameters.      

          b.   For a pump which does not maintain a constant flow
               rate, testing results should be used to generate a
               representative flow profile.  This representative
               flow profile must be included in the device
               labeling.

          c.   Testing should also demonstrate that the device
               can maintain the specified flow characteristics
               despite changes in ambient temperature, pressure,
               or fluid viscosity, which would reasonably be
               expected to be encountered according to the
               intended use of the device.  If significant, the
               effects of these factors should be discussed
               quantitatively in the labeling so the user is made
               aware of the effects of these factors.  

     2.   Comparative Claims

          Additional data may be needed to support comparative
          claims.  

     3.   Unique Designs

          Additional data may be needed to support designs that
          are significantly different from typical designs.

     4.   Biocompatibility Data

          For all fluid path materials certify that the identical
          materials have been used in other legally marketed
          devices under the same use conditions, or provide
          documentation attesting to the biocompatibility of the
          component materials in the finished product according
          to the 1987 Tripartite Biocompatibility Guidance for
          Medical Devices and 1992 draft ISO 194 standard
          (Biological testing of medical and dental materials and
          devices).  With regard to metals, the applicant may
          simply identify device components made of an ASTM grade
          of metal that ASTM certifies is biocompatible, e.g.,
          ASTM 316 stainless steel.   

          Biocompatibility test data may be required for colors
          that are not listed in FDA regulations or are not used
          in other legally marketed devices for a similar
          intended use.
     
     5.   Drug/Biologic and Device Compatibility

          Data demonstrating drug or biologic and material
          compatibility is required if a specific drug or
          biologic is referenced in the device labeling.

     6.   Drug/Biologic Stability

          If the device labeling indicates that the drug or
          biologic is to be stored in the reservoir, stability
          data for the recommended storage period and conditions
          is required.

G.   Software

     The sponsor must indicate the level of concern (see p.16 of the 
document indicated below) for pumps which are software controlled and 
provide appropriate software information according to the "Reviewer 
Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review."

H.   Sterilization Information

     See Attachment 1

I.   SMDA Information

     Summary or Statement of Safety and Effectiveness

     All persons submitting a 510(k) must include either a summary of 
safety and effectiveness information in the 510(k) upon which an 
equivalence determination could be based OR a statement that safety and 
effectiveness information about the [device name] will be made available to 
any interested person upon request.  Safety and effectiveness information 
refers to adverse safety and effectiveness information, descriptive 
information about the new and predicate devices, and performance/clinical 
testing information.

     If the summary option is selected, it should be included on a separate 
page and identified as the Summary of Safety and Effectiveness for [device 
name].

     If the statement option is selected, do not include the word "summary" 
in the statement.

     The content and format of this information is specified in 57FR No. 
82, Tuesday, April 28, 1992, page 18062.

III. COMMENTS

     Address any comments regarding this guidance to:

          Chief, General Hospital Devices Branch
          HFZ-412
          1390 Piccard Drive
          Rockville, MD  20850-4308


Attachments

                               ATTACHMENT 1
                           STERILITY INFORMATION

For a device sold sterile, provide the following information as
detailed in the ODE Blue Book Memorandum #K90-1.

     1.   Sterilization method that will be used.  

     2.   A description of the method that will be used to
          validate the sterilization cycle, but not the
          validation data itself.  Reference to a standard method
          (e.g., AAMI Radiation Standard) usually is sufficient.

     3.   The sterility assurance level (SAL) for the device
          which the firm intends to meet.  An SAL of 10-6 is
          required for devices which contact normally sterile
          areas of the body.

     4.   A description of the packaging to maintain the device's
          sterility (this is not to include packaging integrity
          testing data).

     5.   If sterilization involves EtO, the maximum levels of
          residues of ethylene oxide, ethylene chlorohydrin, and
          ethylene glycol which remain on the device.  The levels
          should be consistent with the draft Federal Register
          Notice on EtO limits.1

     6.   Whether the product is "pyrogen free" and an
          identification of the method used to make that
          determination.2

     7.   The radiation dose, if radiation sterilization will be
          used, and if it has been determined.  Otherwise, amend
          the 510(k) file at FDA when the dose has been
          determined.

References

1.   FDA Proposed Rule, 43 FR 27482 (June 23, 1978), Maximum
     Residue Limits for Ethylene Oxide, Ethylene Chlorohydrin,
     and Ethylene Glycol.

2.   FDA Guidelines on Validation of the Limulus Amebocyte Lysate
     (LAL) Test as an End-Product Endotoxin Test for Human and
     Animal Parenteral Drugs, Biological Products, and Medical
     Devices.

   
ATTACHMENT 2
EXAMPLE OF SIDE BY SIDE COMPARISON TABLE

 ELEMENT OF COMPARISON   SUBJECT 
  DEVICE  
  CLAIMED  
 SE DEVICE  
  #1  
  (CLAIMED 
  SE DEVICE 
 (#2)  
pump type   
intended use(s)   
specific drug.biologic use   
labeling    
components   
pumping mechanism   
administrative sets &
reservoir
   
power requirements   
materials   
flow rates and profiles   
safety/alarm functions   
other capabilities   
SE=Substantially equivalent
The applicant may reference relevant standards in theTable .


ATTACHMENT 3 EXTERNAL INFUSION PUMP REVIEW CHECKLIST 510(k)#:____________ Sponsor:___________ Date:______________ Reviewer:__________ ELEMENT ADEQUATE COMMENTS (e.g., N/A, page #, 30ml, YES NO 18g, PVC, EtO, 10-6, ¾: ") Cover Letter

Labeling
Description of Device
Intended Use(s)
Physical Specifications
Operational Specifications
Biological Specifications
Chemical Specifications
Descriptive Comparison to Legally Marketed Device
Performance Data Supporting Equivalence
Software Information
Sterilization Information
SMDA Information
    

Updated September 5, 1997

horizonal rule

CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page

Center for Devices and Radiological Health / CDRH