Chapter 41. Human Factors and Medical Devices
Harvey J. Murff, M.D.
Harvard Medical School
John W. Gosbee, M.D., M.S.
Department of Veterans Affairs National Center for Patient
Safety
David W. Bates, M.D., M.Sc.
Harvard Medical School
Introduction
Human factors engineering (HFE), also known as usability
engineering or ergonomics, is the study of how humans interact with
machines and complex systems.1 Through the merging of cognitive
psychology, engineering and other disciplines, human factors researchers have
detailed numerous principles concerning device and software program designs that
allow for optimal usage.2 When these principles are violated,
improper use of a machine is more likely to result.3,4 Specific
examples have been detailed during observations of user errors with electronic
infusion devices.5
Medical device misuse is an important cause of medical
error,6,7 and therefore, incorporating human factors methodology into
the design of medical devices has assumed an important role in ensuring patient
safety.2,8 This chapter first describes the use of HFE principles as
a safety practice in the design of medical devices and their evaluation both
prior to and after institutional purchase. Next, medical device alarms and the
contribution of HFE to alarm improvements will be evaluated. Finally, the
chapter reviews the use of preoperative checklist procedures to reduce
anesthesia device failures (see also Chapter 23).
Return to Contents
Subchapter 41.1. The Use of Human Factors in Reducing Device-related Medical Errors
Background
Human factors engineering is a powerful component in the design
of usable, safe medical devices.8 HFE principles can be incorporated
as safety practices that occur at various points during device development and
usage. Industry can use HFE principles at multiple times in the design and
developmental cycle of medical devices and software packages.3
Healthcare institutions can consider results of HFE evaluations when deciding
which products to purchase. Finally, HFE principles can also be incorporated
into the ongoing evaluation of devices that have already been purchased and are
in use. While these practices have high face validity, there has been little
formal study of their effectiveness in reducing medical error. They are
presented here because they may hold promise if scrutinized rigorously, and to
familiarize readers with their potential to reduce medical error.
Design and Developmental Phase
Data collected by the United States Food and Drug
Administration (FDA) in the late 1980s demonstrated that almost half of all
medical device recalls resulted from design flaws.9 In 1990, Congress
passed the Safe Medical Devices Act, giving the FDA the ability to mandate good
manufacturing practices (GMP). These GMP involve design controls for
manufacturers that help ensure the use of HFE within medical device
design.9 As described in the Good Manufacturing Practice Regulation,
Design Control subsection (Title 21-Section 820.30), these include the use of
iterative design and testing during the developmental phase. The Act requires
that designs be "appropriate and address the intended use of the device,
including the needs of the user and patient."10 Multiple human
factors techniques, such as user studies, prototype tests, and task/function
analysis, are utilized in the development and design process.
Manufacturers are required not only to use human factors
principles to repeatedly test the product in all phases of design, but also to
validate the ultimate device design. Validation entails testing the device,
either in an actual clinical situation or a simulation, and documenting that the
device conforms to the individual user's needs. Thus, manufacturers are required
to apply HFE methods through the multiple phases of device design and
development cycles.10
Human factors engineering practices for medical device design
and evaluation have been well described. In 1993 the Association for the
Advancement of Medical Instrumentation and the American National Standards
Institute established guidelines for the incorporation of HFE principles into
medical device design.11 This comprehensive document helped direct
attention to the problem of poor medical device design and helped establish the
design standards necessary to ensure safe medical equipment and thus should
stand as the safety benchmark for industry.
Only limited data are available concerning the application of
HFE principles to medical device design, and most are not published.
Nonetheless, the application of human factors principles during a device's
design phase has been demonstrated to reduce user error. Patient controlled
analgesia (PCA) pumps are a case in point of how HFE principles in product
design reduce user error. User errors associated with poor interface design have
been described with PCA pumps.12,13 Lin and colleagues investigated
whether applying human factors engineering principles to the design of the user
interface of a PCA pump could result in fewer dosage errors as well as less time
spent programming the device.14 Information on device usage was
obtained through cognitive task analysis. This involved observing and
interviewing nurses operating PCA pumps both in the laboratory setting and in
the field. Utilizing this feedback, as well as other human factors design
principles, a "new" PCA pump interface was designed. Twelve recovery room nurses
were required to complete specific tasks with both the standard PCA user
interface and the newly designed interface. There were 29 programming errors on
the traditional interface and 13 on the redesigned interface (an error reduction
of 55%, p<0.01). Furthermore, users were able to program in the necessary
orders in 18% less time.13
Another example involves the design of an ultrasound machine.
In this study, Aucella and colleagues15 interviewed sonographers,
videotaped the ultrasound device being used, and performed usability testing
through simulation to collect information regarding the operator-machine
interface of the ultrasound machine. After their extensive investigations they
implemented over 100 design changes to the console and control panel. Although
errors with the machine were not measured, comments collected by the authors
from the beta operators of the newly designed device suggested that the
resulting machine was much easier to use.
There are enormous numbers of medical devices and software
being designed and developed. Thus the FDA has initiated several regulatory
mechanisms to ensure compliance with these guidelines. Some of the mechanisms
include site inspections of manufacturers, review and approval of medical
devices before marketing, and review of medical device incident
reports.16 Despite the tremendous amount of effort put forth by the
FDA to ensure compliance with the Good Manufacturing Practices Regulation,
individual institutions should critically analyze whether a device they intend
to purchase meets HFE principles for user-centered design.
Device Evaluation Prior to Purchase
Adhering to HFE principles during initial design stages of a
medical device is essential. However, human factors analysis should also be
incorporated into the institutional decision to acquire a new medical device or
software program.3 Device purchasers should strongly consider
institution-specific human factors testing. Usability testing at the
institutional level establishes built-in redundancies to capture any design
problems missed by manufacturers. Furthermore, the users and environments at
individual institutions will differ, possibly in important ways, from the users
and environments in which the device or program was initially designed and
tested. It is important for an institution to be aware of who the intended users
of the device or software will be, as well as where and when they plan to use
the device. The information for such evaluations may be obtained from vendors,
from an in-house analysis, or from independent organizations.
Vendors must be able to prove to the FDA that the user will be
able to operate the medical device in the way in which it was
intended.10 As companies are required to collect human factors
analysis data, it is important that institutions wishing to purchase a new
medical device or software receive and carefully review this information. Gosbee
provides a list of questions to ask a vendor before a purchase, which include:
"How long does it take to learn to operate the system? How long does it take to
complete typical set-up tasks? What are the types and frequency of errors that
could happen, and the systems to thwart them?"3
It is also important to consider the environment in which a
device will be used. Idiosyncratic features of the environment, such as
excessive noise or poor lighting, and differences in user skill or acuity due to
fatigue or otherwise, may affect safety and the device's in-house usability.
Some institutions have developed in-house usability labs, in
order to rigorously test any device before purchasing. The Mayo Clinic uses
simulations to test the usability of medical software before
purchasing.17 By carefully measuring user performance with the
software they are able to uncover latent errors in the design. The usability lab
is also able to measure the time necessary to learn to use the new software.
This important information can help predict the device's or software's influence
on workflow as well as its predilection for operator misuse.
Even without sophisticated usability laboratories, an
institution can use basic human factors techniques to evaluate a product before
purchase.3 Powerful techniques such as cognitive walk-through can be
easily utilized at any institution. This involves observing the end-users of a
product interact with the product. As they attempt to use the device, they are
instructed to "think out loud." Careful observation of the user's actions and
comments can identify potential design flaws that might make it difficult to
utilize the device or software.
Independent organizations are another potential source of
information on device safety. Unfortunately, most independent sources do not
make clear to what degree HFE principles were used in product evaluations,
although they do provide some assessment of safety. One such organization is
ECRI (formerly the Emergency Care Research Institute), a nonprofit international
health services research agency. Another is the Institute of Safe Medical
Practices (ISMP). Both release newsletters and publications regarding product
safety. By searching these and similar databases, institutions can gather
additional information concerning product safety prior to purchasing a device.
ERCI also publishes articles specifically geared to the institutions that might
wish to purchase a medical device or software.
Regardless of the level of pre-procurement testing, some unsafe
designs will not be detected until after the product is in use.3
Therefore, it is important for institutions to continuously evaluate these
products to ensure safety.
Ongoing Device Evaluation
Devices and software at greatest risk for user error should be
systematically evaluated. This is particularly important in areas where multiple
devices are used with different interfaces, such as the operating room or the
intensive care units.3 Furthermore, areas where multiple medications
are stored together should be scrutinized for potential latent errors within
device or software user interfaces prior to user errors occurring.
Resources are available that can help direct an institution's
search. Through publications from the FDA, ECRI, ISMP and similar organizations,
medical device problems identified at other institutions can be targeted. Thus
an important safety practice may be using this published information to search
for latent errors within already purchased medical devices and applying this
information toward a directed product evaluation at the local institution.
Another potential safety practice is to educate practitioners
about HFE principles to increase awareness of medical device user
error.3 Several groups, including the American Nurses' Credentialing
Center and the American Society of Health-Systems Pharmacists, recommend
incorporating HFE training within healthcare curricula as a means to reduce
error.18
To create a culture of safety within medicine, practitioners
must couple the ability to identify potential design weaknesses with a change in
the prevailing culture of silence surrounding medical errors. Educational
programs directed at healthcare providers in training should address both of
these important concerns. Curricula for teaching medical student and medical
residents HFE principles have been described18 and will likely be
adopted at other institutions. Casarett and Helms caution that an unintended
result of error curriculum19 may be that residents become too willing
to attribute an error to system causes. Their concern is that the resident will
ignore any possible individual contribution to the adverse medical event and not
learn from analyses of the event. This concern has been discounted by Gosbee,
stating that any error-in-medicine curriculum should aim to "teach residents to
see when errors are due to inadequate skills and knowledge versus when they are
due to inherent cognitive limitations and biases."18
Return to Contents
Subchapter 41.2. Refining the Performance of Medical
Device Alarms
Background
Numerous aspects of patient care compete for providers'
attention and can reduce their vigilance in monitoring medical devices. Alarms
can alert providers to urgent situations that might have been missed due to
other distractions and have become a necessary part of patient monitoring. In a
study looking at critical incidents within a neonatal intensive care unit, 10%
were detected through alarms.20
However, fundamental flaws in the design of current alarm
systems likely decrease their impact.21 There are reports documenting
some alarm failings in the medical literature,22 but few data address
interventions to improve alarm system effectiveness. For an alarm to be
effective it requires that a medical problem trigger the alarm, that personnel
identify the source and reason for the alarm, and that the medical problem be
corrected prior to patient injury. This section reviews 2 aspects of alarm
safety: (1) the use of HFE principles in the redesign of medical alarms to
improve identification of the source and reason for alarm, and (2) practices in
both device design and programming that may improve safety by decreasing false
positive alarms.
Identification of Alarm Source and Reason
The recognition accuracy of alarms within the operating room is quite low.
When presented with alarm sounds and asked to identify the source,
anesthesiologists, operating room technicians, and operating room nurses
correctly identify the device producing the alarm only 33 to 53.8% of the
time.23-25 Furthermore, experiments suggest that humans have
difficulty reliably recognizing more than 6 alarms at one time.26 The
sheer number of different medical devices with alarms can make it difficult to
discern one alarm from another and studies within the human factors literature
have documented the inability of medical providers to discern between high
priority and low priority alarms.27 While this is a known problem in
operating rooms and intensive care units, how well alarms are recognized in
other settings has not been described.
Some effort has been made to improve alarm systems through
redesign.28 One non-medical study examined ways to improve the
recognition of auditory alarms by comparing abstract alarm sounds with specially
designed alarms using speech and auditory icons.29 Other studies
within the human factors literature have revealed certain acoustical properties
that are more likely to result in a higher sense of perceived urgency by the
operator.
In a series of experiments, Edworthy required subjects to rank the level of
urgency associated with different alarms.30 The acoustical properties
of the alarms were altered for the different subjects. Level of urgency was then
correlated with a specific alarm sound. After ranking a set of acoustic
parameters based on perceived urgency, the experimenters predicted what urgency
ranking the alarm would receive and played the alarms for a new set of subjects.
The correlation between the subjects' urgency rating and the investigators'
predicted ratings was 93% (p<0.0001). Acoustical properties such as
fundamental frequency, harmonic series, and delayed harmonics all affected the
users perceived urgency.
Another study looked at the redesign of an alarm to improve
detectability within the operating room.31 An alarm that was
spectrally rich, frequency-modulated, and contained small amounts of
interpolated silence was detectable with at least 93% accuracy over background
operating room noise. However, both of these alarm experiments have only been
done in laboratory settings. In addition, Burt and colleagues found that when
subjects were required to urgently perform a task, the prior acoustically
manipulated perception of urgency was ignored in order to attend to the
situational urgency of the task.32 Furthermore, with both alarms and
clinical tasks competing for an operator's attention, the newly designed alarm
might not be as discernible. It has continued to be a challenge to create the
best auditory alarm sound to indicate an emergency.
Visual Interfaces for Alarms
Alarms can also be visual. Some research has been done to improve hemodynamic
monitoring device displays. Responses to abnormal values are delayed when
workload for the anesthesiologist is high,33 prompting interest in
improving current visual displays. Furthermore, the clinical decision process
often rests on the practitioner's interpretation of a patient's hemodynamic
parameters. Thus, it is important that this information be presented in a way
that assists with decision making and minimizes errors of interpretation.
Two observational studies have compared different visual
displays of data to traditional visual monitors.34,35 Each evaluated
errors in performing a designated task as well as response time to completion.
One measured how quickly subjects recognized a change in a
parameter34 and the other measured how long it took for
anesthesiologist to manipulate a set of abnormal parameters to a stable
set.34,35 Both studies used computerized simulations of
anesthesiology cases, with subjects serving as their own controls. In one study,
subjects were required to identify when changes in physiologic parameters
occurred using different visual formats.34 Response time and accuracy
to the simulated cases was compared among a histogram, polygon, and numerical
display. Subject responses were more accurate with the histogram and polygon
displays (p=0.01).
In the other study, 20 anesthesiologists with an average
working experience of 5 years were required to perform specific tasks on an
anesthesia simulator35 (see Chapter 45). The tasks consisted of
returning a set of abnormal hemodynamic parameters to normal using intravenous
medications. A specific time for the completion was determined and this time was
compared among 3 different visual interfaces. Trial time was significantly
shorter with the traditional display (p<0.01), yet there were fewer failed
trials using the other monitor displays (26% with the profilogram display, 11%
with the ecological display, and 42% with the traditional display). The slower
time with the non-traditional displays could have resulted from the subject's
lack of experience with such screens. Nevertheless, the newer interfaces
produced fewer failed attempts at arriving at the appropriate hemodynamic
parameters on the simulator, suggesting that these displays might improve the
clinical decision process.
None of the studies comparing traditional auditory alarms and
visual monitor displays reported any adverse event associated with the newer
technology. However these studies are limited by the artificial nature of the
experiments.29,34,35 Anesthesiologists have many tasks to perform
during anesthesia, often amidst great distraction. Attending to monitors is only
one aspect of their workload. Because these laboratory experiments do not
include all of the different "real world" problems and diversions that an
anesthesiologist might face, it is difficult to generalize them to the
workplace. Also, because this experimental task might be taken out of the
context of caring for a patient in the operating room, the subject might simply
focus on the completion of the experimental task and not consider other tasks
that the anesthesiologist would be required to perform in a real
situation.
Decreasing the Frequency of Alarms
Poorly designed device alarms can create not only problems with alarm
recognition but also frequent false positive alarms. Two observational studies
found that from 72 to 75% of alarms during routine general anesthesia did not
require corrective action.36,37 Another study showed that only 3% of
all auditory alarms during routine anesthesia monitoring represented a patient
risk.38 Providers frequently must interrupt clinical tasks to silence
these false positive alarms. More concerning is the fact that when alarms are
unreliable, they tend to be ignored.21,39 This "cry-wolf" effect is a
significant detriment to the optimal performance of alarm systems and may result
in dire consequences when "true alarms" are ignored.
False alarms can be managed in two ways. Devices can be
designed so that they identify and eliminate false alarms before triggering or
users can manipulate alarm parameters to reduce false alarms. User manipulation
can range from adjusting alarm thresholds40 to even turning the
alarms off.22 There are no data describing how often operators reset
alarm parameters to reduce false positive rates.
Some research has focused on the identification of alarm
parameters that improve or optimize alarm accuracy (i.e., to improve the ratio of
true positives to false positives—the "signal-to-noise" ratio). For example,
Rheineck-Leyssius and Kalkman studied how altering an alarm parameter on a pulse
oximeter would affect the incidence of hypoxemia.40 Consecutive
patients admitted to the recovery room of a regional hospital in the Netherlands
after general or regional anesthesia were randomized to either a lower limit of
SpO2 90% or SpO2 85%. The 2 groups were comparable at
baseline. The outcomes measured were hypoxemia, defined by a pulse oximeter
reading less than or equal to 90% or 85%. The authors were also required to
judge if they believed a signal to be artifact versus a true positive. The
authors were blinded as to which group the subject was randomized to during
artifact assessment and data analysis. The relative risk of having a hypoxic
episode (Sp02£85%) in the group with the
lower alarm limit set at 85% (as compared with those with the lower alarm limit
set at 90%) was 3.10 (95% CI: 1.32-7.28, p<0.001). One weakness of this study
was the lack of a bedside observer to verify the validity of the measurement, so
that it is unclear to what degree measurement bias could have affected the
results. The pulse oximeter was considered the "gold standard" for measuring
hypoxia and thus false positives were calculated based on alarm artifact rates
(outliers, loss of signal). Keeping the lower alarm limit for a pulse oximeter
at 90% did reduce the number of patients with hypoxemia, however it also
increased the false positive rate (33% versus 28%). A higher false positive rate
on an alarm could make it more likely that an operator might disregard the
alarm. The majority of alarms were transient and lasting less than 20 seconds.
The authors also noted a 60% reduction in the number of triggered alarms in the
Sp02 90% group by introducing a "theoretical delay" of 15 seconds
between crossing the alarm threshold and actually triggering the alarm. Other
investigators have documented a 26% reduction in mean alarm rate by increasing
the alarm delay from 5 to 10 seconds.41
Overall, only modest evidence supports the practice of not
lowering pulse oximeter lower alarms limit settings below 90%. This intervention
could reduce hypoxemic events with little added cost. However, there would be an
increased number of false positive alarms, which might affect attendance to the
device. Fortunately, newer technological advances in oximetry appear to reduce
false positives rates and may make this less of a problem. In a study in the
Netherlands, a conventional pulse oximeter was compared with a "third
generation" pulse oximeter equipped with a signal processing technique designed
to reduce false positives.42 This "smart" pulse oximeter applied
signal quality tests, differentially amplified the input signal, and applied
motion-detection testing to the identified pulse. The "smart" pulse-oximeter
only triggered one false positive (an alarm that did not coincide with hypoxia)
and had a relative risk of 0.09 (95% CI: 0.02-0.48) for generating a false
positive alarm when compared with conventional pulse oximetry with a 21-second
delay.
Comment
Observational studies have suggested that current alarm systems could be
improved, but future laboratory studies are needed to determine which properties
of alarm systems are most effective to alert operators. These tests must be
followed by field studies and ultimately with trials looking at actual patient
outcomes to determine the best designs. Information concerning the cost and
feasibility of implementing these changes should also be gathered.
False positive alarms remain a significant problem. Few data
exist on the incidence of resetting alarm parameters or at what parameter values
alarm accuracy is optimized. Advances in alarm technology aimed at reducing
false positives appear a promising alternative to resetting
parameters.