Infrared Blackbody
Spectral Characterization |
![](https://webarchive.library.unt.edu/eot2008/20080916110400im_/http://physics.nist.gov/Divisions/Div844/graphics/Horiz_line_spectrum.jpg) |
|
In order to satisfy the needs with blackbodies
infrared calibration and support of the secondary-level blackbody
calibration laboratories, a preliminary spectral radiance scale
realization has been rundertaken at NIST in the temperature range of
232 °C to 962 °C and spectral range of 2.5 µm to
20 µm. Spectral radiance scale realization process includes the use
of Sn, Zn, Al and Ag fixed-point blackbodies (BB), as well as the
transfer of the spectral radiance scale to transfer standard BBs based
on water, Cs and Na heat pipes. Sources are compared by means of a
Fourier Transform Infrared (FTIR) spectrometer, as well as a set of
filter radiometers and a radiation pyrometer. Scales of spectral
emissivity and radiance temperature of BB can be derived from the
spectral radiance scale. The approach to scale realization isillustrated
in figure below.
![](BB_Scale_Realization.gif)
|
The spectral radiance scale is derived from a set of fixed-point BBs,
the design and characterization of which was the first task. Their
temperatures are derived from ITS-90. Secondly, the transfer standard
pyrometers to maintain and interpolate the temperature scale at
particular wavelengths have been developed. The third task was the
construction of a spectral comparator consisting of a spectrometer and
fore-optics for spectral scale transfer. Finally, a set of variable
temperature blackbodies to maintain the scale and interpolate over the
temperature range has been built.
![](FPBBs.jpg)
A set of fixed-point BBs.
For highest accuracy, variable temperature blackbodies
are calibrated in spectral radiance against a pair of fixed-point
blackbodies with interchangeable crucibles of In, Sn, and Zn, and Al, Ag,
and Cu, respectively. Fixed-point blackbodies spectral emissivity also
needs to be accurately characterized. We employ a multi-prong approach:
(1) Monte Carlo ray-trace modeling and calculations,
(2) hemispherical reflectance measurements of the crucible cavity
material flat sample, as well as the cavity itself,
(3) direct spectral emittance measurements of the same samples using
the facility, and (4) comparison of the fixed point blackbodies with
each other as well as with variable temperature heat pipe blackbodies,
using filter radiometers and the facility’s Fourier transform
spectrometer.
|
![](VTBBs.jpg)
A set of variable-temperature BBs |
The Monte Carlo code is used to predict the cavity
emissivity with input of the cavity shape and the emissivity and
specularity of the cavity material. The reflectance measurements
provide emissivity data of both the material and the cavity at room
temperature. The results are used to compare with and validate the
code results. The direct emittance measurements of the material
provide the temperature dependence of the material emittance as code
input. The code predicted results for the cavities at their operating
temperature (freeze points) is then compared with the relative
spectral radiance measurements. Use of this complete set of evaluation
tools enables us to obtain the spectral emissivity of the blackbodies
with reliably determined uncertainties.
The spectral comparator contains fore-optics and a
spectrometer. The fore-optics contains two off-axis aspherical mirrors
and has a low level of out-of-field scatter in the visible and IR.
Medium resolution Fourier Transform spectrometer BOMEM DA3 is used for
relative measurements of spectral radiance.
Two transfer standard pyrometers, RT900 and RT1550,
with spectral responsivity centered at 900 nm and 1550 nm,
respectively, are used to interpolate the temperature scale between
fixed point temperatures, as well as to measure radiance temperature
of the customer blackbody.
|
![](Si_TSP.jpg)
RT900 Si Transfer Standard Pyrometer |
![](InGaAs_TSP.jpg)
RT1550 InGaAs Transfer Standard Pyrometer |
![](BB_chart.gif)
IR Spectral Radiance and Blackbody Spectral Emissivity Scale Realization
|
Customer BB source calibration procedure
- Horizontal/vertical uniformity scan using Transfer Standard
Pyrometer
- Short term temporal stability measurement with the Transfer
Standard Pyrometer.
- Radiance temperature measurements at the BB center with the
Transfer Standard Pyrometer for each temperature set points of
interest.
- Spectral radiance measurements at each temperature set point of
interest with FTIR comparator
- Effective emissivity calculation from measured spectral
radiance, using either set point or Pyrometer radiation temperature
as a reference temperature
|
![](BB_FTIR_EE.jpg)
Spectral Radiance Scale Transfer Example:
Al Fixed Point to Na heat-pipe Variable Temperature Blackbody |
In the spectral band of 8 µm to 14 µm,
the standard deviation of the mean for spectral radiance was typically
at the level of 0.1%. In the spectral band from 3 µm to 5 µm,
scatter of the results for most temperatures is substantially larger,
possibly relating to use of a wide band pyroelectric detector, and
needs further attention. The results show no systematic spectral
features for the transfer standard blackbodies but reveal some
systematic features for one of the customer BBs. The results will be
used to optimize the design of the dedicated Advanced Infrared
Radiometry and Imaging (AIRI) Facility at NIST. |
Publications:
IR spectral
characterization of customer blackbody sources: first calibration
results,
S. Mekhontsev, M. Noorma, A. Prokhorov, and L. Hanssen,
Thermosense XXVIII, ed. by Jonathan J. Miles, G. Raymond Peacock,
and Kathryn M. Knettel, Proc. of SPIE 6205, 620503 (2006).
Emissivity evaluation of
fixed-point blackbodies,
S. Mekhontsev, V. Khromchenko, A. Prokhorov, and L. Hanssen,
9th International Symposium on Temperature and Thermal Measurements
in Industry and Science (TEMPMEKO 2004), June 22-25, 2004, Dubrovnik,
Croatia, Proceedings, Vol. 1, ed. by D. Zvizdic (2004),
pp. 581-586.
|
|